| Catalog Number |
SMS002 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
Fistula (1862); Hemorrhage/Blood Loss/Bleeding (1888); Ulcer (2274); Fungal Infection (2419)
|
| Date of Event |
05/05/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that the dignishield was used on a 43 year old immunocompromised female that was admitted through the emergency department at hospital.The device was left in the rectum after visual evidence of blood in the collection bag.The presence of blood was mentioned to the intensive care unit staff, and it was dismissed as being the patient's menstrual cycle.Menstrual blood was not evacuated through the rectum.The use of this device resulted in irreparable damage to the patient's rectum in the form of ulcers and fistulas.These fistulas became connected to the patient vagina, resulting in fungal infections of both candida and cryptococcus neoformans.The latter fungal infection played an integral role in the patient's death.
|
| |
|
Event or Problem Description
|
|
It was reported that the dignishield was used on a 43 year old immunocompromised female that was admitted through the emergency department at hospital.The device was left in the rectum after visual evidence of blood in the collection bag.The presence of blood was mentioned to the intensive care unit staff, and it was dismissed as being the patient's menstrual cycle.Menstrual blood was not evacuated through the rectum.The use of this device resulted in irreparable damage to the patient's rectum in the form of ulcers and fistulas.These fistulas became connected to the patient's vagina, resulting in fungal infections of both candida and cryptococcus neoformans.The latter fungal infection played an integral role in the patient's death.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed use related.No physical sample was returned for evaluation; therefore, the investigation was based on information provided in the entry description.The available information indicates that the observed rectal bleeding was not investigated and managed as directed in the ifu.The afmea was reviewed for this investigation, and the reported event is addressed.This occurrence does not introduce a new hazard or harm.It represents a known failure mode that has been assessed in the product risk management file in accordance with applicable regulatory standards and internal procedures.The associated risks continue to be monitored as part of ongoing post-market activities in compliance with both regulatory requirements and local procedures.Product labeling was reviewed for this investigation, and the reported event is addressed within the labeling.Please see below instructions to help with continued use of the dignishield stool management system.Warnings:
rectal bleeding should be investigated to ensure no evidence of pressure necrosis from the device.Discontinuation of use is recommended if pressure necrosis is evident.Precautions:
notify a physician if any of the following occur:
persistent rectal pain,
rectal bleeding.Caution should be exercised in using this device in patients who have a tendency to bleed from either anticoagulant / antiplatelet therapy or underlying disease.Potential adverse events:
as with the use of any rectal device, the following adverse events may occur:
pressure necrosis of rectal or anal mucosa,
rectal bleeding,
ulceration of rectal/anal mucosa.Instructions for use:
3.Preparation of patient:
a.The preferred patient position for catheter insertion is the left lateral knee-chest position, although the patient¿s clinical situation may dictate the use of an alternate position.The goal of patient positioning is to maximize sphincter relaxation to ease catheter insertion.A dhr could not be performed since no lot number was provided.Based on the results of the investigation, no evidence was identified to suggest a product quality deficiency contributed to the reported event; therefore, no additional actions are required at this time.Correction: f, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|