An initial event notification was received by sequel med tech, llc, on 28-mar-2026 and forwarded to millyard advanced medical products, llc, on the same day.The user reported that they received a pump error from their twiist automated insulin delivery (aid) system.The user's glucose at the time of the alarm was 384 mg/dl.The user reverted to a backup insulin therapy while awaiting a replacement twiist pump.
|
|
Initial event information was received by sequel med tech, llc, on 28-mar-2026 and forwarded to millyard advanced medical products, llc, on the same day.Based on the information available at that time, the event was initially assessed as non-reportable.Product was later returned to millyard advanced medical products, llc, for investigation on 08-may-2026 and a reportable malfunction was discovered by way of a technical investigation completed on 24-jul-2026.Therefore, for mdr purposes, the date received by the manufacturer (g3) is 24-jul-2026, and the report source (g2) is listed as company representative.Review of the system logs confirmed the reported pump error alarm.Upon disassembly of returned twiist pump, serial number (b)(6), a mechanical malfunction was confirmed, consistent with the reported pump error alarm.A specific cause for the observed malfunction could not be conclusively determined.The contract designer was notified of the investigation findings and will continue to monitor the issue.Additionally, the user's blood glucose level at the time of the pump error alarm was reported as 384 mg/dl.While a causal relationship between the confirmed device malfunction and the reported hyperglycemic event could not be definitively established, this medical device report conservatively considers the hyperglycemia to be an adverse event.-----
the twiist automated insulin delivery system user guide provides information regarding system alarms and the recommended actions associated with them, as well as potential causes of hyperglycemia and ways to prevent it.Users are also instructed to have a backup insulin therapy available at all times.No additional information relevant to the reported event has been made available to millyard advanced medical products, llc, for further investigation.
|