| Model Number |
9-ASD-018 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, an 18mm amplatzer septal occluder was chosen for implant using a 9f amplatzer trevisio intravascular delivery system.Ino issue was noted with the device during the procedure.It was further reported that approximately 24 hours later, the implanter performed a bubble study, which was positive.On (b)(6) 2026, post-procedure, the device was retrieved percutaneously.No adverse events, clinical signs, or symptoms were reported during or after the procedure or retrieval.The implanter considered the event to be related to the procedure rather than the performance of the device.The patient was reported to be stable.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, an 18mm amplatzer septal occluder was chosen for implant using a 9f amplatzer trevisio intravascular delivery system.No issue was noted with the device during the procedure.Approximately 24 hours post-implant, a bubble study was performed and demonstrated a large residual shunt at inferior side of the septum, with bubbles continuing to cross the defect.Based solely on these findings, the physician elected to retrieve the device.On (b)(6) 2026, the device was successfully retrieved percutaneously.No adverse events, clinical signs, or symptoms were reported during or after the implant procedure or retrieval.The physician indicated that the event was related to the procedure rather than device performance.The patient was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of residual shunt post implant was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.The cause for the reported residual shunt could not be determined.It is possible due to procedural condition; however, this cannot be confirmed.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Codes removed:
health effect-clinical code: 4580.
|
| |
|