On (b)(6) 2026, a 64 year old female underwent trusculpt id treatment to the bilateral bra-fat and mid-back areas.Return pads were reported to have been placed on the upper back.Following treatment, a blister was noted on the right back at the edge of a return pad electrode.The physician reported applying a hydrocolloid dressing, providing antibiotic cream and rgn 6 cream, and performing debridement.As of (b)(6) 2026, the affected area was reported to be still healing.No device malfunction or abnormal operation was reported.The cause of the event cannot be definitively determined.The separation between the nearest treatment handpiece and return pad was documented as "a couple of inches".The instructions for use specify a minimum 4 inch (10 cm) separation when treating near the return pads.The reported separation did not meet this requirement.The mid-back treatment is outside of the identified indication for the 2 mhz setting with the 40 cm2 handpiece.
|