On (b)(6) 2026, a 61 year old female underwent trusculpt id treatment to the abdomen using six handpieces.The patient reportedly did not experience pain or excessive heat during treatment.A burn to the lower abdomen was subsequently reported.The available documentation contains an inconsistency regarding when the patient first reported the event to the clinic.One record states that the patient notified the clinic on (b)(6) 2026; however, the clinic had reported the event to cutera on (b)(6) 2026.Therefore, the date the patient first notified the clinic cannot be confirmed.A photograph of the wound that was provided for review is labeled (b)(6) 2026; however, the date the photograph was taken, the accuracy of the displayed date, and the identity of the person who took the photograph have not been confirmed.The wound in the photo appears aged and to involve full-thickness tissue injury.The patient reportedly underwent wound debridement and treatment with hydrocolloid dressings.As of (b)(6) 2026, the area was reported to be still healing.No device malfunction or abnormal operation was reported in association with the treatment.Preventive maintenance was performed on (b)(6) 2026, 49 days, or 7 weeks, before the (b)(6) 2026 treatment.At that time, the system was documented as operating within factory specifications.Based on the available information, the cause of the reported burn cannot be determined.Wounds and burns are identified in the instructions for use as possible adverse effects.
|