• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Fortify Assura VR ICD, US; Implantable Cardioverter Defibrillator (Non-Crt)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Fortify Assura VR ICD, US; Implantable Cardioverter Defibrillator (Non-Crt) Back to Search Results
Model Number CD1357-40Q
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fever (1858); Unspecified Infection (1930)
Date of Event 08/12/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was admitted to the hospital with a fever and signs of infection.Vegetation was noted on the right ventricular (rv) lead via diagnostic imaging.It was also observed that the rv lead exhibited noise oversensing, resulting in pacing inhibition.The implantable cardioverter defibrillator (icd) and the rv lead were explanted on (b)(6) 2026.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Fortify Assura VR ICD, US
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50383208
Report Number2017865-2026-17694
Device Sequence Number13208883
Product Code LWS
UDI-Device Identifier05414734507974
UDI-Public(01)05414734507974
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberCD1357-40Q
Device Catalogue NumberCD1357-40Q
Device Lot NumberA000074466
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-