| Model Number |
1001 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was implanted with the vivistim system on (b)(6) 2024.On (b)(6) 2026, patient indicated to mobia personnel that the device had been explanted approximately one year ago.Follow-up with the implanting site indicated date of explant was on (b)(6) 2025.Patient denied any concerns that the device wasn't working and denied any discomfort when using the device.Patient did not complete the therapy protocol due to non-compliance and discharge by the clinic.No mobia personnel were present at the explant and no explanted device components returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information is expected.Manufacturing records support the device was manufactured and tested to specification.Device was programmed to typical settings during therapy protocol, patient perception level was noted to be relatively high at 2.9ma and well above typical device setting of 0.8ma.
|
| |
|