| Model Number |
0292 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) defibrillation lead fracture was suspected due to a sudden spike in shock impedances to greater than 200 ohms.Subsequently, this rv lead was explanted and replaced.During extraction, it was noted that it was not possible to cannulate the lead past the shocking coil to the tip.No additional adverse patient effects were reported.Product return was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information regarding event cause/resolution and product return status.At this time, no further information is available.Should additional information become available this report will be updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|