Datascope Corp. CARDIOSAVE HYBRID, TYPE G PLUG; System, Balloon, Intra-Aortic And Control
|
Back to Search Results |
|
| Model Number |
0998-00-0800-52 |
| Medical Device Problem Codes |
Display Difficult to Read (1181); Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event site postal code: (b)(6).Updated fields: h11, h2, g6, h6, a3, a4, a2, a1, b5
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that during clinical use, the cardiosave intra-aortic balloon pump (iabp) battery was unable to charge and had a not clear display.Display screen has glitch (red color) on waveform display.Patient was involved.No harm is reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that bay2 slot cannot perform battery charging process.Power management board was re-seated and issue remain same.Full charged battery was returned back.Power management board was replaced the next day and bay2 was able to charge battery.Restarted machine 3 times and display working ok, no glitch issue happened on display screen.Performed display test and result passed.Performed machine simulation for 2 hours and display screen working ok.End user has to monitor the issue for four weeks.Product passed all functional and safety tests per manufacturer specifications.
|
| |
|