| Model Number |
NG-NV-6-30 |
| Medical Device Problem Codes |
Premature Activation (1484); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During the same procedure, two neuroguard devices malfunctioned.A separate medical device report is being made for each device.This report is for the second one.See report 3011471056-2026-00060 for the first.It was further reported that a second device (ng-nv-6-30 neuroguard stent iep system) was attempted through the same 7fr non-contego catheter as the first device and the same resistance and inability to advance occurred, with the system not advancing past the distal end of the 7f catheter.The resistance was reported to have caused the system wire to buckle through the sheath retraction deployment knob.Resistance in the catheter caused an unsheathe and stent deployment inside the guide catheter.Both the 7fr non-contego catheter and neuroguard were again successfully removed together without patient harm.A 6fr non-contego (penumbra benchmark) catheter and a 7x30 non-contego (cordis precise pro rx) stent were used to complete the procedure without issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The event investigation is still ongoing.A supplemental report will be filed when additional information is available.
|
| |
|