The manufacturer received information from a work order on an everflo 120v opi domestic device.Per work order received, the device was alarming, no loss of power alarm, noisy compressor, low purity of 20.9%, cases "melted", and cracked flow control.The problem was able to be duplicated due to faulty board causing audible loss of power alarm, faulty sieve assembly causing low purity and alarming, faulty compressor specs of 12lpm causing low purity and alarming, and foam blocked exhaust port causing overheating and melted cases.The device has not yet been returned to the manufacturer for evaluation.If any additional information is received, a follow up report will be filed.
|