Model Number CI512
Medical Device Problem Codes Failure to Deliver Shock/Stimulation (1133); Unexpected Therapeutic Results (1631); Expulsion (2933)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Per the clinic, the patient had occluding earwax in the ear canal and was seen for earwax removal.During the procedure, an extrusion of the electrode array into the ear canal was discovered.The patient reportedly experienced sounds like rushing water and subsequently lost all sound perception from the implanted device.The device was explanted on (b)(6) 2026, and the patient was reimplanted with a new device during the same surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Nucleus CI512 Cochlear Implant with Contour Advance Electrode
Common Device Name
Nucleus 24 Cochlear Implant System
Manufacturer (Section D)
Cochlear Ltd
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer (Section G)
Cochlear Ltd
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
amirah rosly
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key50399728
Report Number6000034-2026-02971
Device Sequence Number15746349
Product Code MCM
UDI-Device Identifier09321502014564
UDI-Public(01)09321502014564(11)151002(17)171001
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P970051
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Type of Report Initial
Report Date (Section B) 08/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/01/2017
Device Model NumberCI512
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/02/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-