Medtronic Ireland ENDURANT IIS BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
|
Back to Search Results |
|
| Catalog Number |
ESBF3614C103E |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An endurant iis stent graft (b)(6) was implanted during the endovascular treatment of a 58.5 mm abdominal aortic aneurysm.It was reported that during the index procedure, after visual inspection prior to use, the stent graft was delivered through the left common femoral artery to the implantation site.During deployment of the device, the physician applied gentle backward force pulling, rather than advancing, to keep the radiopaque markers aligned with the renal artery lines drawn on the screen.No resistance was felt by the physician.The device was delivered easily into position and implanted successfully.During removal of the delivery system, which was performed in accordance with the instructions for use (ifu), difficulty was encountered while attempting to remarry the delivery system , as the grey handle would not slide fully to the blue handle after advancing approximately halfway.Manual rotation of the blue handle was required to bring it into contact with the grey handle.Following removal, inspection identified a large twist or kink in the distal portion of the graft cover portion of the delivery device.The delivery device was successfully removed from the left groin, and the procedure was completed.Per the physician, the cause could not be determined; however, a tight aortic bifurcation was noted as a possible contributing anatomical factor.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis conclusion the reported difficulty in removing the delivery system and kink icould not be confirmed based on the pre-implant films provided; therefore, the cause of the event cannot be determined.Procedural angiograms showing stent graft deployment and removal attempts were not available for assessment.While the cause remains indeterminate, moderate calcification in the common iliac arteries may have been a contributing factor.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|