(b)(4).Date sent 8/24/2026.D4 batch #
photo analysis:
this is an analysis of photo submitted for evaluation.During the visual analysis, the following was observed:
the photo shows a device with a reload loaded, all drivers up and the knife exposed, also the staple height selector can be seen on the green color.Please note is possible that the knife exposed noted on the reload is consist with the firing knob was not returned completely prior to opening the device causing that the knife not returned completely to its home position.Based on the photo the event described is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.A manufacturing record evaluation was performed for the finished device lot number of 960d88 / 03sr75, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|