It was reported that, during a pancreaticoduodenectomy, the device could be used without issues at the first firing during the initial duodenal transection, but the firing knob became stiff and did not move at the second firing during the gastric body transection.The doctor tried to advance the knife several times, but it did not move and the knife did not reach the stomach body.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
(b)(4).Date sent: 8/24/2026.D4 batch # 757d69.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ntlc75 device was received with no apparent damage and with a reload present.The reload was received with the proximal 3 driver up and the swing tab was noted to be broken, making the reload nonfunctional.In addition, the reload was also damaged in the reload shroud at the same side where the remaining portion of the swing tab is placed.The device was tested for functionality with a test reload and it fired, cut, and formed all the staples as intended.The staple line and cut line were complete and the staples meet the staple form release criteria.It is possible that the damage on the swing tab was due to the reload was not properly loaded into the device, causing the damage to the swing tab and not allowing the device to fire.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meets the required specifications prior to shipment.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 757d69, and no non-conformances related to the reported complaint condition were identified.
|