| Model Number |
SGC0801 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available information was investigated, and the reported split sgc soft tip was confirmed as a deformed soft tip via device analysis.The reported failure to advance could not be replicated in a testing environment.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, the cause of the reported failure to advance sgc was unable to be determined.The reported kinked sgc tip was likely a result of procedural circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4 degenerative mitral regurgitation (mr) for a mitraclip procedure.During the procedure, a steerable guide catheter (sgc) was attempted to be advanced within the patient whne excessive resistance was identified.Troubleshooting was performed and the device was removed from the patient.The device was examined and it was visualized that there was a split in the tip of the guide.A replacement sgc was selected and advanced within the patient.The procedure was performed successfully and an mitraclip was implanted.The procedure was discontinued, the final mr was reduced to grade 1.There were no adverse patient effects or clinically significant reported.
|
| |
|
Search Alerts/Recalls
|