• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Laboratories Alinity c Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Laboratories Alinity c Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE Back to Search Results
Catalog Number 03R67-01
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed depressed sodium when processing on alinity c processing module.Sid (b)(6)(male 81 years old) generated sodium of 118 mmol/l that repeated 141 mmol/l.The repeated result matched patient historical result and was reported out of lab.The customer normal range: sodium 135-145 mmol/l.No impact to patient management was reported.
 
Additional Manufacturer Narrative
The field service representative (fsr) performed quarterly maintenance on the alinity c processing module, serial number (b)(6), found the external waste tubing was obstructed, causing back pressure at the hcw pump, and cleaned the lines.However, a definitive part was not identified as cause of the issue, therefore, the alinity c, serial number (b)(6) was deemed the cause.The module function was verified with a control run.The service history of the alinity c, serial number (b)(6) returned no additional complaint tickets for the module generating discrepant sodium patient results or message code 3062.A review of tracking and trending did not identify any trends for the alinity c processing module with regards to the customer reported event.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.Labeling was reviewed and was found to address the customer reported issue.Based on our investigation, no systemic issue or deficiency was identified with the alinity c processing module, serial number (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Alinity c Processing Module
Common Device Name
ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Manufacturer (Section D)
Abbott Laboratories
1915 hurd drive
irving TX 75038
Manufacturer (Section G)
Abbott Laboratories
1915 hurd drive
irving TX 75038
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
+49()6122582960
MDR Report Key50408415
Report Number3016438761-2026-00464
Device Sequence Number15746424
Product Code JJE
UDI-Device Identifier00380740137380
UDI-Public(01)00380740137380
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K170316
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03R67-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/24/2026
Initial Report FDA Received Date08/24/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
Patient SexMale
-
-