(b)(4)
date sent: 10/02/2026
d4: batch #911d96
investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one vasecr35 reload was received unfired, with no apparent damage, and with an opened package.The returned reload was loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the reload performed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 911d96, and no non-conformances were identified.
|