• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Fortify Assura DR ICD, US; Pulse Generator, Permanent, Implantable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Fortify Assura DR ICD, US; Pulse Generator, Permanent, Implantable Back to Search Results
Model Number CD2357-40Q
Medical Device Problem Code Failure to Convert Rhythm (1540)
Health Effect - Clinical Code Arrhythmia (1721)
Date of Event 08/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient presented remotely via (b)(6).Review of the transmission revealed that the patient's implantable cardioverter defibrillator (icd) exhibited inadequate anti-tachycardia pacing (atp) which failed to convert the ventricular tachycardia (vt), and the patient was still in vt.Programming changes were discussed, no changes or interventions were reported.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Fortify Assura DR ICD, US
Common Device Name
Pulse Generator, Permanent, Implantable
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50410548
Report Number2017865-2026-17768
Device Sequence Number21125364
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public(01)05414734508094(10)S000082287(17)230331
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberCD2357-40Q
Device Catalogue NumberCD2357-40Q
Device Lot NumberS000082287
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/27/2026
Initial Report FDA Received Date08/24/2026
Supplement Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Durata; Tendril
Patient Age65 YR
Patient SexMale
-
-