This additional supplemental report is being submitted to provide the results of the final investigation.Updated field(s): h2, h4, h6: b17 to b01, d15 to d02
the device was returned to olympus for inspection, and the reported failure was¿confirmed.A definitive root cause could not be identified.Based on the results of the investigation, there was shredding and torn pieces of sheath but found that the needle could be extended and retracted smoothly.The most probable cause of the complaint is cause traced to component failure, likely due to stress of repeated use, external factors, or handling.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|