• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Gyrus ACMI, Inc. Single Use Aspiration Needle NA-U401SX; Biopsy Needle

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Gyrus ACMI, Inc. Single Use Aspiration Needle NA-U401SX; Biopsy Needle Back to Search Results
Model Number NA-U401SX-4021-A
Medical Device Problem Codes Material Frayed (1262); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
No additional information was received from the customer.
 
Additional Manufacturer Narrative
This additional supplemental report is being submitted to provide the results of the final investigation.Updated field(s): h2, h4, h6: b17 to b01, d15 to d02 the device was returned to olympus for inspection, and the reported failure was¿confirmed.A definitive root cause could not be identified.Based on the results of the investigation, there was shredding and torn pieces of sheath but found that the needle could be extended and retracted smoothly.The most probable cause of the complaint is cause traced to component failure, likely due to stress of repeated use, external factors, or handling.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Single Use Aspiration Needle NA-U401SX
Common Device Name
Biopsy Needle
Manufacturer (Section D)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
Gyrus ACMI, Inc.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50410655
Report Number3011050570-2026-00840
Device Sequence Number18797047
Product Code FCG
UDI-Device Identifier04953170409974
UDI-Public04953170409974
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160098
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNA-U401SX-4021-A
Device Lot NumberPW311575
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/02/2026
Initial Report FDA Received Date08/24/2026
Supplement Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-