| Model Number |
D-EVOLUTFX-2329 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the attempted implant of a transcatheter aortic valve, upon the first deployment attempt, the valve was unable to be deployed.Subsequently, the valve was recaptured into the delivery catheter system (dcs).Upon the second deployment attempt, the valve remained unable to be deployed.Subsequently, the valve was recaptured, and the system was withdrawn from the patient.A pre-implant balloon aortic valvuloplasty was performed, and a second valve was opened and used.Per the physician, the patient's heavily calcified anatomy contributed to the inability to fully deploy the valve.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5.The event description was corrected.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|