Synthes GmbH 3.5/4.5MM VA PPFX PROX FEMUR PLATE/8HOLES/RIGHT/STERILE; Plate, fixation, bone
|
Back to Search Results |
|
| Catalog Number |
02.221.114S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Subject id: (b)(6)
study no: no information provided.Clinical adverse event received for fixation failure (displacement of fixation) with specifically: displacement of screws, displacement of plate, breakage of screw(s)
device related: yes
procedure related: yes
date of event: 29-jun-2026.Date of implant: (b)(6)2025.Date of revision: (b)(6) 2026.Device location: distal femur.Treatment/impact: infection at the time of the periprosthetic plate placement (2025 (b)(6)), resulted in medical or surgical intervention: yes, required in-patient hospitalization or prolongation of existing hospitalization: yes, resulted in a permanent impairment of a body structure or function: yes, 2-stage surgery - osme (which means osteosynthesis material extraction--meaning, removal of the failed plate and screws) + explantation tkr + spacer on (b)(6) 2026, second stage surgery planned for (b)(6) 2026.Depuy synthes products used:
catalog number: no information provided.Lot number: no information provided.Component type: plate
description: no information provided.Catalog number: no information provided.Lot number: no information provided.Component type: screw
description: no information provided.Catalog number: no information provided.Lot number: no information provided.Component type: screw
description: no information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|