| Model Number |
470183-19 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the permanent cautery hook started to smoke at the tip of the instrument.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the da vinci product with an alleged issue to perform failure analysis.Failure analysis could not replicate nor confirm the customer complaint regarding the item started smoking at the tip of the instrument.The permanent cautery hook (pch) was tested on an in-house system.The instrument was tested on an in-house system, and it passed the recognition and engagement tests.The pch instrument moved intuitively with a full range of motion in all directions.The pch instrument was found to be fully functional.
|
| |
|