| Model Number |
VTICM5 12.6 (-13.00/1.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation.Lens explanted.Cause of the event was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Lens rotation and secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|