| Model Number |
5101 |
| Medical Device Problem Codes |
High impedance (1291); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with a rechargeable snm ipg was experiencing pain at their battery site.The patient also noted the stimulation was inadequate.A red light was later noticed on the device; however, an impedance check was not performed.The device was explanted and the patient is stable with no reported complications.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a rechargeable snm ipg was experiencing pain at their battery site.The patient also noted the stimulation was inadequate.A red light was later noticed on the device; however, an impedance check was not performed.The device was explanted and the patient is stable with no reported complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
product code: additional product code qon.Block g4:
premarket / 510(k) #: additional premarket/510(k) p190006.Block d10:
concomitant product:
model number/catalog number: 1201
serial number: (b)(6)
model/catalog description: tined lead
unique identifier (udi) #: (b)(4).Block h11:
the device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.Good faith effort attempts were made to try and retrieve the device, however the device was retained by the explanting provider.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the events of "message - error codes displayed on rc", "implant pain" and "impedance high" were defined in the product risk documentation.These event types have been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
product code: additional product code qon.Block g4:
premarket / 510(k) #: additional premarket/510(k) p190006.Block d10:
concomitant product:
model number/catalog number: 1201,
serial number: (b)(6),
model/catalog description: tined lead,
unique identifier (udi) #: (b)(4).
|
| |
|