Model Number 5101
Medical Device Problem Codes High impedance (1291); Communication or Transmission Problem (2896)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 07/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with a rechargeable snm ipg was experiencing pain at their battery site.The patient also noted the stimulation was inadequate.A red light was later noticed on the device; however, an impedance check was not performed.The device was explanted and the patient is stable with no reported complications.
 
Event or Problem Description
It was reported that a patient with a rechargeable snm ipg was experiencing pain at their battery site.The patient also noted the stimulation was inadequate.A red light was later noticed on the device; however, an impedance check was not performed.The device was explanted and the patient is stable with no reported complications.
 
Additional Manufacturer Narrative
Block d2b: product code: additional product code qon.Block g4: premarket / 510(k) #: additional premarket/510(k) p190006.Block d10: concomitant product: model number/catalog number: 1201 serial number: (b)(6) model/catalog description: tined lead unique identifier (udi) #: (b)(4).Block h11: the device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.Good faith effort attempts were made to try and retrieve the device, however the device was retained by the explanting provider.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the events of "message - error codes displayed on rc", "implant pain" and "impedance high" were defined in the product risk documentation.These event types have been accounted for during analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Additional Manufacturer Narrative
Block d2b: product code: additional product code qon.Block g4: premarket / 510(k) #: additional premarket/510(k) p190006.Block d10: concomitant product: model number/catalog number: 1201, serial number: (b)(6), model/catalog description: tined lead, unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Axonics
Common Device Name
Stimulator, Electrical, Implantable, For Incontinence
Manufacturer (Section D)
Axonics Modulation Technologies, Inc.
15515 sand canyon ave
irvine CA 92618
Manufacturer (Section G)
Axonics Inc
15515 sand canyon ave
irvine CA 92618
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50416010
Report Number2124215-2026-45359
Device Sequence Number10213669
Product Code EZW
UDI-Device Identifier10810005340479
UDI-Public(01)10810005340479(11)240316(17)270316
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number5101
Device Catalogue Number5101
Device Lot NumberAX1G198536
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/11/2026
Initial Report FDA Received Date08/24/2026
Supplement Report FDA Received Date10/02/2026
Date Device Manufactured03/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age80 YR
Patient SexUnknown
-
-