It was reported that this implantable cardioverter defibrillator (icd) delivered one burst of anti-tachycardia pacing (atp) and seven shocks as inappropriate therapy for atrial fibrillation (af) with rapid ventricular response (rvr).Technical services (ts) was consulted and discussed stored as well as programming options based on the patients rhythm.Additionally, the patient reported that the device felt warm and hot intermittently.Ts performed an analysis of device battery and it was determined the battery was depleting normally, with power consumption stable.The device function is normal since high power events would be reflected in power consumption or monitoring voltage, with both normal and no device issues noted.This device remains in service.No additional adverse patient effects were reported.
|