• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Stryker GmbH Hoffmann II Connecting Rod Micro D3 x 30mm; Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Stryker GmbH Hoffmann II Connecting Rod Micro D3 x 30mm; Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component Back to Search Results
Catalog Number 50796030
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Hoffmann II Connecting Rod Micro D3 x 30mm
Common Device Name
Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Manufacturer (Section D)
Stryker GmbH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
Stryker GmbH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
+1(201)831-5000
MDR Report Key50418745
Report Number0008031020-2026-01318
Device Sequence Number13426830
Product Code KTT
UDI-Device Identifier04546540389893
UDI-Public04546540389893
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K033480
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Remedial Action Other,Notification
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number50796030
Device Lot Numberunknown
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/27/2026
Initial Report FDA Received Date08/24/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2827-2026
Patient Sequence Number1
Patient SexUnknown
-
-