• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Allergan Pleasanton CoolSculpting System; Dermal cooling pack/vacuum/massager

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Allergan Pleasanton CoolSculpting System; Dermal cooling pack/vacuum/massager Back to Search Results
Catalog Number UNK Control Unit
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperplasia (1906)
Date of Event 07/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reported that they experienced paradoxical hyperplasia to the arms and abdomen after coolsculpting treatments on (b)(6) 2021 and (b)(6) 2021, using an unknown applicator, 1 month after treatment.Patient described the following: "hardening" and "it is bulging".Patient received lymphatic massage.
 
Event or Problem Description
Abbvie received additional information that the patient was treated on (b)(6) 2026 using the cooladvantage+ applicator and presented with paradoxical hyperplasia (ph).
 
Additional Manufacturer Narrative
Paradoxical adipose hyperplasia (ph) is a known potential adverse event addressed in the product labeling.According to the coolsculpting user manual, under rare adverse events, ph is characterized by a visibly enlarged tissue volume within the treatment area, which may develop two to five months after treatment.Surgical intervention may be required.Ph is not related to any coolsculpting device failure mode but it is included in the risk management files of the device because it is a risk that is inherent to the use cryolipolysis for localized fat reduction.A supplemental report will be submitted if and when additional information is obtained.
 
Additional Manufacturer Narrative
Additional information: a3, a4, a5, b3, b5 (treatment date, applicator).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CoolSculpting System
Common Device Name
Dermal cooling pack/vacuum/massager
Manufacturer (Section D)
Allergan Pleasanton
4410 rosewood dr
pleasanton CA 94588
Manufacturer (Section G)
Allergan (Pleasanton)
4410 rosewood dr
pleasanton CA 94588
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key50419013
Report Number3007215625-2026-00428
Device Sequence Number18806503
Product Code OOK
UDI-Device Identifier00816417020001
UDI-Public(01)00816417020001
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberUNK Control Unit
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/25/2026
Initial Report FDA Received Date08/24/2026
Supplement Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight152 LB
-
-