| Model Number |
U125 |
| Medical Device Problem Codes |
Extracardiac stimulation (1463); Incorrect, Inadequate or Imprecise Result or Readings (1535); Battery Problem (2885)
|
| Health Effect - Clinical Codes |
Device Overstimulation of Tissue (1991); Discomfort (2330)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) reverted to safety mode.When attempting a software update, the device went from ten months remaining to seven months remaining then reverted to safety mode."0xa" and "oscillator mismatch" codes were also recorded.The patient stated they felt uncomfortable and experienced stimulation while in office.Technical services (ts) recommended device replacement.This crt-p was explanted, replaced, and returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|