Model Number FPRPR3508
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft frost occurred.An icerod plus 90 degree needle was selected for use.During device preparation, the needle frosted up the shaft.The needle was exchanged and the procedure was completed.There were no patient consequences as a result of this event.
 
Additional Manufacturer Narrative
D3/g1: 2069203 investigation results: device history record (dhr) review: the device history record review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause traced to component failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Icerod Plus 90 Degree Needle
Common Device Name
Unit, Cryosurgical, Accessories
Manufacturer (Section D)
BTG Yokneam
tavor building 1
industrial park
yokneam
IS 
Manufacturer (Section G)
Boston Scientific Corporation
tavor building 1
industrial park
yokneam
IS  
Manufacturer Contact
ricardo amaro castro
4100 hamline ave n
arden hills, MN 55112
+1(952)930-6396
MDR Report Key50419161
Report Number2124215-2026-45419
Device Sequence Number18669635
Product Code GEH
UDI-Device Identifier07290104830563
UDI-Public(01)07290104830563(17)280322(10)38908211
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110946
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFPRPR3508
Device Catalogue NumberFPRPR3508
Device Lot Number0038908211
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/24/2026
Date Device Manufactured03/23/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight158 LB
-
-