Axonics Modulation Technologies, Inc. Axonics; Stimulator, Electrical, Implantable, For Incontinence
|
Back to Search Results |
|
| Model Number |
1201 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that upon connecting the patient's neurostimulator (ins) to the clinician programmer, an error message appeared and open impedances were noted.The patient stated that their device was not functioning, but the patient later noted that they were fine with no pain or discomfort.The patient underwent an x-ray.A revision was performed and a new lead was placed successfully.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b:
product code: additional product code qon.Block g4:
premarket / 510(k) #: additional premarket/510(k) p190006.Block h11:
the device was not returned for analysis; therefore, a technical analysis could not be performed.A media inspection was performed, in the picture provided the impedances are open.The issue was confirmed.It was confirmed the device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Search Alerts/Recalls
|
|
|