| Model Number |
9799 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the device is displaying inaccurate hr readings.There was no patient impact or consequence reported.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device is displaying inaccurate hr readings.There was no patient impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: masimo has reached out to the customer to request the return of the device.The device has not been returned to masimo to allow an analysis to be performed. if the device is returned for evaluation or new information is obtained, a follow up report will be submitted.Health effect impact code: no adverse event; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: the returned device was evaluated.The device was found to have a broken and recessed cable connector frame on the instrument housing.This damage was not found to correlate with the customer's reported issue nor impact the device's performance related to accuracy during the investigation.The device was able to obtain measurements for all enabled parameters throughout the duration of testing, and all alarm conditions were audible and visual.No product performance issues related to spo2 or pr were identified in both the manual and simulator testing.The customer's reported issue was not replicated, root cause cannot be identified.H6.Component code - connector frame: health effect impact code: no adverse event; no patient impact.
|
| |
|