| Model Number |
D121 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) delivered potential inappropriate therapy for an atrial arrythmia.Technical services (ts) reviewed the stored episode and showed atrial fibrillation (af) with rapid ventricular response (rvr) rhythm that exceeded 200 beats per minute detected in the ventricular fibrillation (vf) zone.Anti-tachycardia pacing (atp) bursts and electric shocks were delivered for the episode.The health care professional (hcp) reported that the patient felt the delivered shock.The device was operating as programmed.No adverse patient effects were reported.This icd remains in service.
|
| |
|