William A. Cook Australia, PTY LTD Zenith Branch Endovascular Graft Iliac Bifurcation; MIH System, Endovascular Graft, Aortic Aneurysm Treatment
|
Back to Search Results |
|
| Model Number |
G60345 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The implant procedure was an iliac branch with a zbis on the left, with competitor grafts for everything else: evar hypo stent with vbx, gore limbs to bridge.They sealed into the top of the zbis with the 14.5 ispi leg of the exculder.Brisk endoleak was visible at level of the zbis flow divider.They then relined the vbx thinking maybe it was a type c and occuleded above the flow divider with a coda.The leak was still there.They then relined the ipsi limb with another limb.Leak was still there.Reballooned and took different angles to rule out type 2 or any other leak elsewhere.Doctor seems confident now after reflection it could only be coming from the zbis at the area of the flow divider somehow.(the physician called the cook representative to complain that she is fairly certain now that this is type 4 endoleak of some kind due to the graft.).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in progress.Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|
|
|