Medtronic Ireland ENDURANT IIS BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
|
Back to Search Results |
|
| Catalog Number |
ESBF3214C103E |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the index procedure to treat a 55 mm abdominal aortic aneurysm, an endurant iis stent graft delivery system kinked at the mid-shaft during implantation.The aortic diameter at the renal artery was 27 mm, the proximal aortic neck length was 35 mm, and the neck had 10 degrees of angulation.The iliac anatomy was calcific and tortuous, and force was required to deliver the system.The physician was not comfortable continuing to use the kinked delivery system.The original device was not implanted, and was replaced with a new device.The replacement device was delivered without difficulty using a non-medtronic 20 fr sheath.Theevent was resolved.The endurant iis was prepared and used in accordance with the instructions for use (ifu), and no damage was noted to the device prior to insertion into the patient.According to the physician, the cause of the event was anatomy-related due to calcific and tortuous vessels.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|