The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.Due to discoloration of light guide (lg) lens, illumination is uneven, which leads to poor quality image, and due to slipping-out of light guide bundle (lg-bundle), no illumination is obtained, which leads to poor quality image.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|