| Model Number |
1054260 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information that a trilogy 100 ventilator was returned to a third-party service center for evaluation.There was no patient involvement, and no harm or injury reported.On device evaluation, a ventilator service required alarm was noted in the device log indicating an internal battery fault.This is the final power source so it may impact therapy.No device repair was completed; the device was processed as scrap and discarded per local regulations.Additional findings not related to the malfunction/issue: the sd card slot was noted to be damaged, and the speakers were not plugged in.
|
| |
|
Additional Manufacturer Narrative
|
|
The device mentioned in this complaint has exceeded its useful life per the applicable ifu.Therefore, the dhr for this device will not be reviewed at this time.However, the dhr files for this device are available and can be accessed upon specific request.The reported failure of the internal lithium-ion battery is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Search Alerts/Recalls
|
|
|