The provided photo confirmed the reported balloon rupture; however, there was no evidence of blood on the device to indicate that it had been used on the patient.While the balloon rupture was confirmed, the exact root cause of the reported issue could not be determined.Balloon rupture is an inherent risk associated with the use of this product.Balloon rupture is an inherent risk of using this product.As stated in the ifu: avoid extended or excessive exposure to fluorescent light, heat, sunlight, or chemical fumes to reduce balloon degradation.Do not exceed the recommended maximum balloon liquid capacity (see specifications).Due to the increased rate of occurrence of this issue, capa 2026-007 was created to document and investigate the failure.The production and traceability record for the device was reviewed; no issues were found during manufacturing or packaging that would cause or contribute to the reported event.All quality control tests were completed successfully and met specification.
|