• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Endo-Surgery, LLC. N/A; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Endo-Surgery, LLC. N/A; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number CDH25A
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 10/6/2026 d4 batch # 169d37.Investigation summary the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual examination of the returned sample found no apparent damage to the cdh25a device.The anvil with the washer uncut were received in a plastic bag.The breakaway washer was received uncut, and no staples were present in the device.As the returned device contained no staples and anvils are interchangeable within the same product model, it is possible that the returned anvil with uncut washer does not belong to the returned device.The device was reloaded with staples and it was tested for functionality with the returned washer and anvil.The device fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 169d37, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50429216
Report Number3005075853-2026-07062
Device Sequence Number13380821
Product Code GDW
UDI-Device Identifier10705036003458
UDI-Public(01)10705036003458(17)290930(10)326D38
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K983536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial,Followup
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCDH25A
Device Lot Number326D38
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/25/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/21/2026
Supplement Date Received by Manufacturer09/25/2026
Initial Report FDA Received Date08/25/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-