(b)(4).Date sent 10/6/2026
d4 batch # 169d37.Investigation summary
the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual examination of the returned sample found no apparent damage to the cdh25a device.The anvil with the washer uncut were received in a plastic bag.The breakaway washer was received uncut, and no staples were present in the device.As the returned device contained no staples and anvils are interchangeable within the same product model, it is possible that the returned anvil with uncut washer does not belong to the returned device.The device was reloaded with staples and it was tested for functionality with the returned washer and anvil.The device fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 169d37, and no non-conformances were identified.
|