| Catalog Number |
N180802 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during thrombectomy for a left distal external iliac vein femoral deep vein thrombosis, the nitrex guidewire was inserted through a 4 fr micropuncture introducer sheath and repositioned several times while attempting to access a vein containing thrombus.The distal end of the wire separated from the wire body within the vessel.The detached distal wire component was retrieved using a snare.The procedure was continued with a new device.The event was associated with an approximately 20-minute surgical delay.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|