It was reported that the device had over infusion of midazolam.Infusion was ordered to infuse at a rate of 9ml/hr for a total volume of 57ml.A "local unbranded 60ml syringe" was used.There was patient involvement, but harm is unknown.Additional details added from the incident report received on 20-aug-2026."on the day of incident, the patient was appeared to be struggling with the mechanical ventilator at ~11pm despite being prescribed high dose of sedation.Upon assessment, i checked the loaded midazolam syringe in alasis pump & observed that syringe contained volume of air, instead of the medication solution.There was no alarm even.Upon confirmation from evening staff nurse, she changed new loaded midazolam 50ml syringe at 09:00pm."
"patient was sedated with midazolam and morphine on full mode mechanical ventilation, despite of high dose sedation she was resisting ventilation.".
|
|
The event logs have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.Per 803.52(f)(11)(iii) the information provided represents all of the known information at this time.The complainant or reporter was unable or unwilling to provide any further patient, product, or procedural details to the manufacturer.
|