| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The parent reported scalp swelling and redness following head impact in may this year.Hearing performance with the ci deteriorated significantly.The user has been re-implanted.
|
| |
|
Event or Problem Description
|
|
The parent reported scalp swelling and redness following head impact in may this year.Hearing performance with the ci deteriorated significantly.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.Other mechanical damages found are attributable to the removal surgery.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|