Model Number 2104
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Bacterial Infection (1735)
Date of Event 07/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A barostim system was implanted on (b)(6) 2023 and an ipg replacement occurred on (b)(6) 2026.It was noted that the incision looked good at discharge and the patient confirmed during a 24-hour check call that everything looked okay.On (b)(6) 2026, during the patient's one week follow up, the pocket looked infected, and the patient was advised to go to the emergency room.The incision was red, draining, and part of the incision had been dehisced.Cultures were taken and the result was positive for staphylococcus epidermidis.The patient was administered oral antibiotics.A revision was done, and the pocket was debrided, cultures were taken and it was closed.The root cause of the infection was not disclosed by the physician.As of (b)(6) 2026, the patient continued to report symptoms of infection.On (b)(6) 2026, the device was explanted.
 
Additional Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.The device history and sterilization record for this device serial number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.Bioburden, pyrogen, lal, and cleanroom cae testing results were below control limits for the quarter the ipg was manufactured.The device met material, assembly, and quality control requirements.Cvrx id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Barostim Neo2
Common Device Name
Implantable Pulse Generator
Manufacturer (Section D)
CVRx, Inc
9201west broadway avenue
suite 650
minneapolis MN 55445
Manufacturer (Section G)
CVRx, Inc
9201west broadway avenue
suite 650
minneapolis MN 55445
Manufacturer Contact
vincent mbibi
9201west broadway avenue
suite 650
minneapolis, MN 55445
MDR Report Key50431751
Report Number3007972010-2026-00099
Device Sequence Number9479664
Product Code DSR
UDI-Device Identifier00859144004623
UDI-Public(01)00859144004623(17)271031
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P180050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number2104
Device Catalogue Number100065-202
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/18/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2026
Initial Report FDA Received Date08/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
Patient SexMale