Catalog Number PVE35A
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unspecified surgical procedure, the device was fired the first time perfectly and second time it fired but they then couldn't release the stapler.They had to use manual override to then release the jaw from the tissue.There was no patient consequence nor surgical delay reported.No additional information provided.
 
Additional Manufacturer Narrative
(b)(4) date sent: 8/25/2026.D4: batch #886d47.Additional information was requested, and the following was obtained: 1.Could you please clarify if there were any patient consequences? no consequences but it caused stress to the surgeon and the theatre team and they had to open a new device once they removed this from the patient after using the manual override.2.Could you please clarify if there were any change in the post-operative care of the patient as a result of the event? none that i know of.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pve35a device was returned with no reload present.During visual inspection, the bailout door was noted to be out of position; additionally, the bailout lever was damaged, likely due to interaction with the cam bailout.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.It is possible that an outside the normal use force was applied on the lever bailout as it was pushed down once the device was bailout.The event described of difficult to open could not be confirmed as the device opened and closed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 886d47, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50432262
Report Number3005075853-2026-07076
Device Sequence Number9480152
Product Code GDW
UDI-Device Identifier10705036014584
UDI-Public(01)10705036014584(17)281231(10)886D47
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K141952
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPVE35A
Device Lot Number886D47
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/11/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date08/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-