date sent: 8/25/2026.D4: batch #886d47.Additional information was requested, and the following was obtained:
1.Could you please clarify if there were any patient consequences?
no consequences but it caused stress to the surgeon and the theatre team and they had to open a new device once they removed this from the patient after using the manual override.2.Could you please clarify if there were any change in the post-operative care of the patient as a result of the event? none that i know of.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pve35a device was returned with no reload present.During visual inspection, the bailout door was noted to be out of position; additionally, the bailout lever was damaged, likely due to interaction with the cam bailout.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.It is possible that an outside the normal use force was applied on the lever bailout as it was pushed down once the device was bailout.The event described of difficult to open could not be confirmed as the device opened and closed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 886d47, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|