(b)(4).Date sent: 8/25/2026.D4: batch#: 629d43.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee45a device was returned with no apparent damage and with no reload present.During the functional test, the sound of a motor was heard, and the knife didn't move forward to the lockout position, this is consistent with a device being bailout.The device was disassembled to verify the condition of the internal components and the lever was damaged near interaction with the cam bailout.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 629d43, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device psee45a lot number a9882a and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|