An impella cp was inserted percutaneously via right femoral artery access site in a 64-year-old male patient for acute myocardial infarction/ cardiogenic shock.The patient¿s comorbidities were coronary artery disease, renal insufficiency, peripheral artery disease (pad) and peripheral vascular disease (pvd).The patient¿s clinical state prior to initiation of support was society for cardiovascular angiography intervention (scai) shock stage scai shock stage e identified as the body fails to respond to maximal lifesaving therapies.The patient was on inotropes, vasopressors and respiratory support.The patient was on concurrent extracorporeal membrane oxygenation (ecmo) as the primary hemodynamic support device with the impella in place to vent the left ventricle.During/after introducer insertion, the patient developed right lower limb ischemia, diagnosed by loss of pulses in the affected extremity.Due to severe coagulopathy and active bleeding concerns, including reported act >800 seconds, inr 17, and hemoglobin 5.8 g/dl, the implanting physician elected not to peel away the 14 french introducer because of concerns for worsening bleeding in a heavily calcified vessel.Attempts to obtain antegrade access for limb perfusion were unsuccessful.Following transfer, the receiving physician also elected not to remove the sheath due to ongoing bleeding concerns and worsening overall clinical status.It was reported that if the patient stabilized, the medical team would subsequently address the ischemic limb; however, the ischemia did not resolve.The impella device remained in place and was not removed due to the event.The patient expired less than 24 hours after support initiation while still receiving support.Leaving the 14 french sheath in place is not recommended, as it may restrict the critical blood flow to the lower extremities.When combining emco and impella supports this significantly increases the bleeding risks due to mandatory systemic anticoagulation, large-bore arterial and venous cannulation.The death is multifactorial as the patient required the need for multiple forms of mechanical circulatory support, including va-ecmo and impella cp support.Contributing risk factors include severe pad/pvd, calcified vasculature, profound coagulopathy, and active bleeding.The patient continued to require escalating vasopressor and inotropic support and progressed into scai stage e identified as the body fails to respond to maximal lifesaving therapies.
|
|
A4 is unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc or its employees that the report constitutes an admission that the product, abiomed inc, or its employees, caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|