Medtronic Navigation, Inc. AQUAMANTYS 6.0 BIPOLAR SEALER; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
|
Back to Search Results |
|
| Model Number |
23-112-1 |
| Medical Device Problem Codes |
Smoking (1585); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a handpiece.It was reported that during an above-the-knee amputation procedure, a generator produced a loud noise during activation (other case).The generator was switched out intraoperatively, and a similar loud noise was observed with the replacement generator during activation while using the same device.A loud pump noise was also reported to be always present when the handpiece was activated.The generator was then reported to have started ¿smoking¿ while the button was pressed, and use of the device was discontinued intraoperatively after the smoking was observed.The machines were changed during the case and the device was re-primed, and the procedure was completed without further use of the device.Patient was present and unknown patient impact.Additional information was received.It was reported that the generator would not pump fluid.The cause of the noise was the bad tubing of the handpiece.The handpiece was not stopped immediately after the noise was heard; use continued until it started smoking.The noise continued during use and activation, not just during initial priming.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3): no parts have been received by the manufacturer for evaluation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|