MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
| Model Number |
8637-40 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Codes |
Itching Sensation (1943); Muscular Rigidity (1968); Irritability (2421); Insufficient Information (4580)
|
| Date of Event |
12/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from multiple sources (healthcare provider, clinical study) regarding a patient who was receiving gablofen via an implantable pump.It was reported that the patient experienced agitation, itching in groin and increased tone.Baclofen withdrawal suspected.The patient was treated with oral baclofen and oral valium, 65 mcg single bolus of baclofen given via pump.The patient responded to bolus with decreased tone, decreased agitation, decreased clonus, and decreased itching.The baclofen dose was serially increased from 271.5 mcg/day to 622 mcg/day over a 2 year period.On (b)(6) 2022 a dye study was completed in preparation for pump replacement.During replacement surgery tissue noted around pump connector, requiring the spinal portion to be cut from pump.After replacement of spinal segment, slow egress of cerebrospinal fluid (csf) noted, a kink in the catheter was identified and resolved with repositioning and anchoring of catheter connector.Csf flow improved post revision compared to pre-revision.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 8596sc, serial# (b)(6), implanted: (b)(6) 2012.Product type catheter section d information references the main component of the system.Other relevant device(s) are: product id: 8596sc, serial/lot #: (b)(6), ubd: 02-may-2014, udi#: (b)(4).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|