| Catalog Number |
32000014 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility reported that the vent connection tubing on the venous side ruptured and caused blood leakage.The physician observed the leak 10 minutes following support start.There was no indication of patient serious injury.The complaint sample was discarded and unavailable for product investigation.
|
| |
|
Event or Problem Description
|
|
A user facility reported that the vent connection tubing on the venous side ruptured and caused blood leakage.The physician observed the leak 10 minutes following extracorporeal membrane oxygenation (ecmo) support start.The patient's cannulas were clamped with hemostats and the kit was exchanged.There was no indication of patient serious injury.The complaint sample was discarded and unavailable for product investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d4 plant investigation: nonconformity was not observed during the manufacturing process.No similar complaints have been reported about this batch number.The complaint sample was not available for evaluation.The provided photos and video show a crack and blood leak at the luer-lock negative adapter of the venous venting line.Cracks in this component could be due to a problem with the injection molding process.The supplier of the component has implemented corrective measures in the injection molding process.
|
| |
|