| Model Number |
CD3369-40Q |
| Medical Device Problem Code |
Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the non-implanted implantable cardioverter defibrillator (icd) was found to be operating in the backup mode.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint of unexpected reset was confirmed.The device was received at backup mode of operation, which was discovered while device was still in the box.Analysis of device image determined backup occurred due to bus error and hardware reset of xena integrated circuit.After reloading device firmware, further environmental testing could not recreate the backup event.Additional testing found the device battery at normal levels.Longevity assessment found the device was operating above the elective replacement indicator (eri), within the expected voltage range.
|
| |
|